Stable and fast rising in temperature with high precision
The column oven with big volume can be installed two columns
Optional cooling function to meet the testing of biochemical samples
|
EX1600CO |
EX1600COⅡ |
Temperature Control Principle |
Air circulation heating + semiconductor temperature control |
Temperature Range |
≥ Room temperature +10℃
≤85℃ |
≥ Room temperature -10℃
≤85℃ |
Temperature Control Range |
4℃-85℃ |
Temperature Control Precision |
±0.1℃ |
Chromatographic Columns Installed |
Two |
Size and Weight |
61 (L) × 62 (W) × 26 (H) cm, about 17kg |
Power Source |
110V/220V |
二.EX1600 Pump
Big cam reciprocating parallel infusion, more stable and reliable
Automatic rinse device selected. Using the buffer system is more handy
Many human-based auxiliary and prompting functions enable users to replace consumables more conveniently
Model |
EX1600HP/EX1600LP |
EX1600PP |
Principle of Liquid Delivery |
Big cam-driven, dual-piston reciprocating parallel pump |
Max Working Pressure |
42Mpa |
20Mpa |
Flow Range |
0.001mL/mim~9.999mL/min |
0.01mL/min~40.00mL/min |
Flow Precision |
RSD<0.06% |
RSD《1% |
Pressure Fluctuation |
0.1Mpa |
0.2Mpa |
GLP Function |
Pressure curve, flow curve, recording frequency of piston seal used |
Package Size and Weight |
61(L)×44(W)×35(H) cm, about 17kg |
Power Source |
110V/220V |
三. EX1600 AS
The operation and control of EX1600AS autosampler, which can easily implement the whole process from sample injection to analyzing result, is made entirely by software.
The capacity of EX1600AS can be adjusted from 12 vials to 768 wells. 2 Micro well plates according to SBS standards; 96-well high/low and 384-well low formats; 48-vial or 12-vial trays.
Three injection modes: as Full loop, partial loopfill and μL pickup mode.
The application of advanced sampling technology can achieve rapid sampling and cycle cleaning, and efficient needle track cleaning can achieve low residue.
When using perforated plates or sample bottles, such articles may be used openly or sealedly.
Option of sample cooling function.
Sampling Range |
1~5000μL with increment of 1μL (10mL sample loop, optional) |
Syringe Volume |
500μL (2500μL, optional) |
Sample Volume |
2×48 Sample plate (2mL sample vial);(12-hole sample plate, 96-hole and 384-hole plate, optional) |
Switching Time of Sampling Valve |
<100ms |
Location Precision of Sampling Needle |
≤0.6nm |
Sampling Circulation Time |
17s (including 60s for washing needle) |
Sampling Method |
All sample loop injection, partial sample loop injection, and carrying injection |
Sampling Repeatability |
All sample loop method RSD<0.3%; partial sample loop method RSD<0.5%; carrying injection method RSD<1.0% |
Sampling Residue |
<0.05% |
Cooling Range |
4℃ to -3℃ room temperature |
Noise |
<70dB |
四.EX1600 UV/UV-Vis Detector
Using all-digital communication for full-process instruments control through RS232 or USB interface
2000-h high-performance deuterium lamp
Unique design of flow cell with double parallel conical bores, effectively amplifying light intensity and energy
Automatic spectrum scanning
Optional automatic rinse device. When buffer salt is used as mobile phase, it is more convenient and can reduce faults
Model |
EX1600UV |
EX1600UVⅡ |
Light Source |
D2 lamp |
D2 lamp + W lamp |
Wavelength Range |
180nm-700nm |
180nm-900nm |
Wavelength Precision |
±0.1% |
Spectral Bandwidth |
8nm |
Noise |
±0.2×10-5AU |
Drift |
0.3×10-4AU |
Measuring Range |
-2AU~2AU |
Response Time |
0.1,0.2,0.5,1,2, adjustable |
Wavelength Time Sequence |
Optional change of time wavelength |
GLP Function |
Time of using lamp, recording D2 and W light energy |
Size and Weight |
61(L)×44(W)×35(H) cm, about 17kg |
Power Source |
110V/220V |
五.EX1600 RI
EX1600RI differential refractive index detector is of excellent sensitivity, stability and reproducibility, and able to detect the most subtle difference of polymeric physical properties, and the analysis result is highly reliable.
Superior operability:
1. The autostart function can complete the complex operation from the displacement of reference cell solution to verification of stability of the baseline.
2. Affirmation of the help function can determine device formation.
Excellent Stability:
Optimal temperature control reduces the power-on stabilization time, and thus the excellent stability of the baseline is achieved.
Refraction index range |
1.00~1.75 |
Range |
0.2~3.0mL/min |
'Recorder output range selection |
0.25~512µRIU |
noise |
2.5nRIU( response; 1.5Sec) |
Cell volume |
8ul |
Drift |
0.2µRIU/h(water;1mL/min;PURGE OFF) |
Linearity range |
600µRIU |
Bias adjustment range |
10µRIU |
Bias decompound energy |
50nRIU |